Open Access

Table 1.

Specification of the four DOSTEL detectors. The thickness (column 2), area (column 3), mass (column 5), and energy deposition ranges (column 5) are shown. For the telescope detectors these energy ranges can be converted to LET in water.

Detector Thickness/(μm) Area/(cm2) Mass/(g) dE range/(keV)
LET range/(keV μm−1 in water)
Upper tele. detector 154 6.93 0.249 dEi = [28.7,…, 22612]
LETi = [0.084,…, 66.3]
Lower tele. detector 315 6.93 0.509 dEi = [42.5,…, 51248]
LETi = [0.061,…, 73.5]
Pin diodes 300 2.31 0.161 dEi = [44.4,…, 56292]
no pathlength limitation