Issue |
J. Space Weather Space Clim.
Volume 13, 2023
|
|
---|---|---|
Article Number | 10 | |
Number of page(s) | 13 | |
DOI | https://doi.org/10.1051/swsc/2023007 | |
Published online | 26 April 2023 |
Technical Article
The radiation environment over the African continent at aviation altitudes: first results of the RPiRENA-based dosimeter
1
Center for Space Research, North-West University, 2531 Potchefstroom, South Africa
2
Institut für Experimentelle und Angewandte Physik, Christian-Albrechts-Universität zu Kiel, 24118 Kiel, Germany
* Corresponding author: mosothomoshegodfrey@gmail.com
Received:
2
August
2022
Accepted:
2
March
2023
The radiation environment over the African continent, at aviation altitudes, remains mostly uncharacterized and unregulated. In this paper, we present initial measurements made by a newly developed active dosimeter onboard long-haul flights between South Africa and Germany. Based on these initial tests, we believe that this low-cost and open-source dosimeter is suitable for continued operation over the African continent and can provide valuable long-term measurements to test dosimetric models and inform aviation policy.
Key words: Space weather / Aviation dosimetry / Active dosimeters / Absorbed dose rates / Africa
© M. Mosotho et al., Published by EDP Sciences 2023
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
1 Introduction
Primary cosmic rays (CRs), in particular those of galactic and solar origin, are the most dominant particle species interacting with atmospheric molecules creating a hazardous secondary radiation field at, and above, aviation altitudes (Reames, 1999; Vainio et al., 2009; Desai & Giacalone, 2016). The secondary radiation produced at aviation altitudes is heavily dependent on several factors, including the phase of the solar cycle, the geomagnetic cutoff rigidity (Pc) determining which CRs can penetrate the protective magnetosphere, the residual atmosphere about the aircraft, and relevant aircraft shielding (Feynman & Gabriel, 2000; Smart & Shea, 2005; Desmaris, 2016). The International Committee on Radiological Protection (ICRP) has recognized and classified flight personnel’s CR exposure as an occupational hazard. The ICRP guidance recommends that exposure to ionising radiation should be below 20 mSv/year for radiation workers, while 1 mSv/year is recommended for the general public and occasionally exposed commercial flight passengers (ICRP60, 1991; ICRP132, 2016). In Europe and USA studies by e.g. Thierfeldt et al. (2009) and the US Federal Aviation Administration1 have shown that the annual maximum effective dose for regular passengers and pilots are about 6–7 mSv. In Africa, it is unclear what the annual effective dose range for pilots or regular passengers will be at aviation altitudes as no assessment of their radiation exposure has been conducted. Nonetheless, at flight altitudes, radiation exposure during a single flight will normally not cause immediate health effects or exceed any recommended limits. But, for flight personnel, exposure increases their long-term radiation dose accumulation and this might ultimately contribute to an additional health risk. However, it is possible that the minimum exposure recommendation can be exceeded in a single flight during a severe solar particle event (Matthiä et al., 2009; Fujita et al., 2021). On the Earth’s surface, the temporal flux of cascading particles produced within the atmosphere has been monitored for years using a network of ground-based neutron monitors (NMs, e.g. Simpson, 2000; Moraal et al., 2000). With the recent record-breaking CR fluxes observed from NM stations during the previous solar minima in 2009 and 2020 (e.g. Moraal & Stoker, 2010; Oh et al., 2013; Strauss & Potgieter, 2014; Fu et al., 2021), there is additional motivation to routinely monitor radiation levels at aviation altitudes even during very quiet solar periods.
To support efficient and safe air navigation, the International Civil Aviation Organization’s (ICAO’s) revised airspace procedural manual includes expected radiation dose exposure as part of the mandatory information to be provided from designated space weather centres as part of flight planning (ICRP132, 2016; ICAO, 2019). In the African airspace, the South African National Space Agency’s (SANSA’s) space weather center2 provides advisories to the Air Traffic and Navigation Service (ATNS) regarding enhanced radiation levels at flight altitudes for civil aviation3. There is thus a need to characterise and quantify the localized radiation levels on-board commercial aircraft in near real-time using active dosimeters (see Möller, 2013; Ritter et al., 2014).
To meet the requirements for space weather application, a suitable dosimeter must ideally be lightweight and relatively small to make it easy to deploy in different locations on ground level, onboard commercial aircraft flights, or even during stratospheric balloon launches. Apart from being a plug-and-play instrument, it should also have very low power consumption and an environmental sensor to monitor (cabin) pressure and temperature. For this work, the RPiRENA4 (i.e. RPi stands for Raspberry Pi and RENA stands for Read-out Electronics for Nuclear Applications) detector, which is an open-source project (instrument schematics, all software, and calibration techniques are available publicly), was modified to serve as an active dosimeter.
This instrument and its calibration are discussed in more detail in the Appendices A and B. The radiation field at flight altitudes differs from that at ground level, as well as that produced by the calibration sources used here. The detector is calibrated for the low energy range of the secondary CR radiation, while, for the higher energies, we assume a linear extrapolation up to 7 MeV in silicon. This was shown to be a reasonable assumption by Ritter (2013) using the same low energy calibration sources. In the current configuration, the dosimeter includes two silicon PIN diodes that can be operated in single mode, where we obtain the absorbed dose in each diode, or in telescope mode which allows us to determine the linear energy transfer and therefore also the quality factor of the incident radiation. In this work, we present the initial results of this newly developed dosimeter during two long-haul flights from South Africa to Germany.
2 Dosimetric quantities
When incident ionizing radiation interacts with matter, they deposit energy in the material. The energy per unit mass deposited in the target material, by the incident ionising radiation, is a quantity called the absorbed dose. For our silicon diodes (SDs), this quantity is obtained from the energy deposition spectra of each SD as(1)where is the number of events detected in the ith energy bin, with a midpoint located at E i , and M is the SD’s mass. The energy deposition of each particle (or event) is calculated by using the calibrated conversion described in Section A.3 in the Appendix A. Absorbed doses measured using silicon-based detectors are given in units Gy. To indirectly mimic the biological damage in human tissue as a result of exposure to ionising radiation, the absorbed dose measured in silicon has to be converted to a comparable value in water, , with the absorbed dose in the water now given in units of Sv (see Bartlett, 2004; Hellweg & Baumstark-Khan, 2007; ICRP103, 2007; Flowers et al., 2018). We use the standard approach for this conversion,(2)where (Beaujean et al., 2005). A corresponding absorbed dose rate, , can also be calculated by dividing by the time interval of the measurement. With the dosimeter operated in telescope mode, coincidence events (an event detected simultaneously in both diodes) allow us to approximate the linear energy transfer (LET) spectrum of the incident radiation. The ionizing particle moves through one SD, depositing an amount of energy ΔE, after which it is detected in the second diode. The linear transfer, again in silicon, is then approximated as(3)where 〈leff〉 is the effective thickness of the SD, calculated in Section A.5 in the Appendix A. In this approach, the LET is therefore estimated as the energy loss of charged particles interacting with the target material per path length and allows us to approximate the quality factor of the incident radiation and therefore its biological effectiveness. Similar to the absorbed dose, the LET in silicon needs to be converted to LET in water. This is done through the expression(4)where and ρsi are the densities of water and silicon with values of 1 g/cm3 and 2.33 g/cm3, respectively (see Ritter et al., 2014) and, once again, .
3 First results
The dosimeter was placed on two long-haul flights: on 10 June 2019 from Johannesburg (FAJS) to Frankfurt (EDDF) and on 15 June 2019 from Munich (EDDM) to FAJS, each lasting between 11 h and 12 h. These measurements were performed inside a commercial aircraft and serve as a first test of the instrument in a mixed radiation field over the African continent. To interpret radiation dose rate measurements obtained onboard aircraft, it is essential to discuss the possible space weather effects and influences on the measurements. During these flights, the geomagnetic conditions were assessed using the Kp index and in both flights, it was below 3, an indication that the magnetosphere was virtually undisturbed5. Furthermore, there were no solar events during these flights. Thus, the contribution of solar energetic particles can be excluded. The omnipresent GCR component observed from the NMDB6 polar stations shows marginal modulation variation rates below 2% for both flight routes. As a result, changing space weather effects during both flights can be assumed to be negligible.
Figure 1A shows a visualization of the flight routes, while Figure 1B shows the corresponding altitude and effective vertical cutoff rigidity, Pc, during these flights. Data for the flight routes were downloaded from FlightAware7. For these long-haul flights, the Pc varied significantly, covering both the North and Southern hemispheres and crossing the magnetic equator, during each flight. The relevant flight data for flights are summarized in Table 1.
Figure 1 Panel A shows the test flight routes from FAJS to EDDF (in blue) while the return flight to FAJS from EDDM is shown in red. Panel B shows the flight altitudes (in terms of atmospheric depth) as a function of the corresponding vertical geomagnetic cutoff rigidity, Pc. Panel A image credit: Google maps (https://sourceforge.net/projects/py-googlemaps/). Supplementary data are available on the website. |
Long-haul flight information.
Figure 2 shows the count rates from both SDs during both test flights. The panels on the left represent the flight from FAJS to EDDF and the panels on the right represent the flight from EDDM to FAJS. Blue and red curves show the results for both diodes, while the green curves are the count rates of the coincident events. In Figure 2A, during take-off and landing, it can be seen that with increasing altitude (decreased atmospheric shielding) the number of detected particles increases sharply. During the cruising phase the count rate changes due to changing values of Pc (note the decrease towards the equatorial regions) and changes in the flight altitude, especially the noticeable increase towards the end of the first flight.
Figure 2 Panel A shows the measured count rates (SD 1 in blue, SD 2 in red, and coincidence counts in green) during the two tests. Panel B shows the measured absorbed dose rate in each diode, with a minute cadence, while panel C shows the accumulated absorbed dose, . The black curves in the bottom panels show results for the CARI-7A model for 0.3 mm of silicon. The left panels are for the flight from FAJS to EDDF and the right panels are for EDDM to FASJ. Supplementary data are available on the website. |
In Figure 2B, the dependence of the calculated absorbed dose rate in silicon on the altitude and Pc is also visible. At these altitudes, ionizing particles, e.g. muons, electrons, and protons, make up the bulk of the particles contributing to the absorbed dose rate measured in silicon. In Figure 2C, an accumulated absorbed dose () of ~11.5 μGy was recorded on the flight from FAJS to EDDF in both diodes, while a dose of ~11 μGy was recorded on the return flight. The data for the two flights in terms of an accumulated flight dose are also presented in Table 1. The measured absorbed doses were compared to CARI-7A8 model simulations, for model information see e.g. Copeland (2017) while an overview of the model calculations performed (and shown in Figs. 2B and 2C) are given in Appendix B. For both flights, the model predicts higher accumulated absorbed doses, i.e. about a factor of ~1.5 higher, compared to the measurements.
The generated LET spectra for both flights, integrated over the total flight duration, are given in Figure 3. The spectrum is constructed by making use of the calculated geometrical factor of the SDs in telescope mode as discussed in Section A.4 in Appendix A. The LET spectrum of the FAJS to EDDF flight is given in red and the return flight is in blue. Clearly visible is the prominent peak from minimal ionizing charged particles and the steep decrease towards higher energy depositions following a power law. Figure 3 also shows, in orange, the quality factor from ICRP103 (2007). We can combine the LET spectra, and the dependence of the quality factor on the LET, to determine the mean quality factor as(5)
Figure 3 LET spectra accumulated for the flight to EDDF (in red) and a return flight to FAJS (in blue). The quality factor, as defined by ICRP103 (2007), is shown in orange. Supplementary data are available on the website. |
As we have no significant LET events with keV/μm, we find . The calculated mean quality factor does, however, deviate strongly for individual flights since it depends strongly on the few high LET events that are observed during the rather short measurement intervals with a dosimeter with a small detection area (see Ritter, 2013; Ritter et al., 2014). Using the calculated , the dose equivalent (in water) can be calculated from the calculated absorbed dose in water as(6)
4 Discussion and conclusion
Much research has been devoted to studying the effects and possible impacts of radiation exposure during commercial flights. ICAO’s recent amendments specifically focus on the short-term impacts and effects of radiation exposure on avionics and the health and safety of flight personnel. As such, near real-time and reliable estimates of the radiation levels at aviation altitudes are essential. In this article, an assembled, tested, and calibrated dosimeter, based on the open-source RPiRENA detector, is described. Initial long-haul test flights of this prototype are very promising and present the first attempt at characterizing the radiation levels at aviation altitudes over the African continent.
We believe that the RPiRENA-based dosimeter described in this work offers the opportunity to expand pre-existing coverage of the existing dosimeter network, especially over Africa. Measurements from this instrument are and will continue to be, of high importance to establish, enrich, and maintain a dialogue regarding radiation dosimetry and radiation protection between general members of the public, academia, and the private sector, especially within African airspace, where there is little or no data available.
Since the prototype discussed in this article was tested, the instrument has been developed further, including the addition of a GPS antenna to directly log the position (when such a measurement is possible) of the instrument and refining the software to be more user-friendly. One drawback of the current design and this is, in general, true for all active dosimeters using SDs, is the inability to correctly account for the absorbed dose due to neutrons, especially at aviation altitudes. Due to the SD-based detector’s small interaction cross section with neutrons (Ritter, 2013; Ritter et al., 2014), the SDs used in this work have a low efficiency for neutron detection. At cruising altitudes, it is believed that the radiation absorbed dose is dominated by the absorbed dose received from neutrons, generated as a result of the interactions of primary CRs with atmospheric molecules (Bartlett et al., 2002; Poje et al., 2015; Meier et al., 2020) and subsequently with the aircraft components (Knipp, 2017). Work is, however, underway locally in South Africa to account for the neutral component at flight altitudes through the development of a dedicated neutron detector (Buffler et al., 2015) that can be flown and operated independently alongside the silicon-based dosimeter discussed here in order to measure the dominant components of the complex radiation field at aviation altitudes.
Supplementary material
Supplementary material provide by the authors. Access here
Acknowledgments
The author acknowledges financial support from the South African National Space Agency’s (SANSA, Space Science Division) PhD bursary program and the National Research Foundation (NRF) of South Africa’s Professional Development Programme (PDP). The NRF’s unique identifier number is 104898 – S&F – Scholarships & Fellowships Programme. Opinions expressed and conclusions arrived at are those of the authors and are not necessarily to be attributed to SANSA or the NRF. We acknowledge the support provided by the North-West University (NWU) Instrument Making Department. Figures in this paper were prepared with Matplotlib (Hunter, 2007). The editor thanks Kyle Copeland and an anonymous reviewer for their assistance in evaluating this paper.
See a detailed description of the project at http://www.ieap.uni-kiel.de/et/people/stephan/rpirena/
See data from omniweb: https://omniweb.gsfc.nasa.gov/form/dx1.html
The Neutron Monitor Database (NMDB): http://www.nmdb.eu/
References
- Bartlett D. 2004. Radiation protection aspects of the cosmic radiation exposure of aircraft crew. Radiat Prot Dosimetry 109: 349–355. https://doi.org/10.1093/rpd/nch311. [CrossRef] [Google Scholar]
- Bartlett D, Tanner R, Hager L. 2002. High energy neutron response characteristics of a passive survey instrument for the determination of cosmic radiation fields in aircraft. Radiat Prot Dosimetry 100: 519–524. https://doi.org/10.1093/oxfordjournals.rpd.a005927. [CrossRef] [Google Scholar]
- Beaujean R, Burmeister S, Petersen F, Reitz G. 2005. Radiation exposure measurement onboard civil aircraft. Radiat Prot Dosimetry 116: 312–315. https://doi.org/10.1093/rpd/nci095. [CrossRef] [Google Scholar]
- Böttcher SI. 2018. Calibration of HET/EPT Flight Models using data obtained in thermal Vacuum. Tech Rep 6951. http://www.ieap.uni-kiel.de/et/people/stephan/solo/eda/gse/hetept/tvcal/hetept-tvcal.pdf. [Google Scholar]
- Buffler A, Comrie AC, Smit FD, Wörtche HJ. 2015. Neutron spectrometry with EJ299-33 plastic scintillator for En = 10–100 MeV. IEEE Trans Nucl Sci 62(3): 1422–1428. https://doi.org/10.1109/TNS.2015.2430455. [CrossRef] [Google Scholar]
- Copeland K. 2017. CARI-7A: Development And validation. Radiat Prot Dosimetry 175(4): 419–431. https://doi.org/10.1093/rpd/ncw369. [Google Scholar]
- Desai M, Giacalone J. 2016. Large gradual solar energetic particle events. Living Rev Sol Phys 13: 3. https://doi.org/10.1007/s41116-016-0002-5. [CrossRef] [Google Scholar]
- Desmaris G. 2016. Cosmic radiation in aviation: radiological protection of Air France aircraft crew. Ann ICRP 45(1 suppl): 64–74. https://doi.org/10.1177/0146645316636009. [CrossRef] [Google Scholar]
- Feynman J, Gabriel S. 2000. On space weather consequences and predictions. J Geophys Res 105: 10543–10564. https://doi.org/10.1029/1999JA000141. [CrossRef] [Google Scholar]
- Flowers P, Theopold K, Langley R, Robinson WR, Bott S, et al. 2018. Chemistry. OpenStax, Rice University, Houston, Texas. https://openstax.org/details/books/chemistry. [Google Scholar]
- Fu S, Zhang X, Zhao L, Li Y. 2021. Variations of the galactic cosmic rays in the recent solar cycles. Astrophys J Suppl Ser 254(2): 37. https://doi.org/10.3847/1538-4365/abf936. [CrossRef] [Google Scholar]
- Fujita M, Sato T, Saito S, Yamashiki Y. 2021. Probabilistic risk assessment of solar particle events considering the cost of countermeasures to reduce the aviation radiation dose. Sci Rep 11: 17091. https://doi.org/10.1038/s41598-021-95235-9. [CrossRef] [Google Scholar]
- Hellweg C, Baumstark-Khan C. 2007. Getting ready for the manned mission to Mars: The astronauts’ risk from space radiation. Die Naturwissenschaften 94: 517–526. https://doi.org/10.1007/s00114-006-0204-0. [CrossRef] [Google Scholar]
- Hunter JD. 2007. Matplotlib: A 2D graphics environment. Comput Sci Eng 9(3): 90–95. https://doi.org/10.1109/MCSE.2007.55. [CrossRef] [Google Scholar]
- ICAO. 2019. Manual on space weather information in support of international air navigation (Doc 10100), Tech Rep. https://store.icao.int/en/shop-by-areas/safety/meteorology. [Google Scholar]
- ICRP60. 1991. 1990 Recommendations of the International Commission on Radiological Protection. Ann ICRP 21: 1–3. https://www.icrp.org/page.asp?id=5. [CrossRef] [Google Scholar]
- ICRP103. 2007. The 2007 recommendations of the International Commission on Radiological Protection. Ann ICRP 37: 2–4. https://www.icrp.org/page.asp?id=5. [Google Scholar]
- ICRP132. 2016. Radiological protection from cosmic radiation in aviation. Ann ICRP 45(1): 1–48. https://www.icrp.org/page.asp?id=5 https://www.icrp.org/page.asp?id=5. [CrossRef] [Google Scholar]
- Knipp DJ. 2017. Essential science for understanding risks from radiation for airline passengers and crews. Space Weather 15(4): 549–552. https://doi.org/10.1002/2017SW001639. [CrossRef] [Google Scholar]
- Labrenz J. 2014. Radiation field in low earth orbit: measurements from MATROSHKA DOSTEL onboard ISS, Ph.D. thesis, Faculty of Mathematics and Natural Sciences. https://macau.uni-kiel.de/receive/diss_mods_00014952. [Google Scholar]
- Leo WR. 1987. Techniques for nuclear and particle physics experiments: A how-to approach. Springer-Verlag. https://books.google.co.za/books?id=0sPvAAAAMAAJ. [Google Scholar]
- Matthiä D, Heber B, Reitz G, Meier M, Sihver L, Berger T, Herbst K. 2009. Temporal and spatial evolution of the solar energetic particle event on 20 January 2005 and resulting radiation doses in aviation. J Geophys Res 114: A08104.https://doi.org/10.1029/2009JA014125. [Google Scholar]
- Meier MM, Copeland K, Klöble KEJ, Matthiä D, Plettenberg MC, Schennetten K, Wirtz M, Hellweg CE. 2020. Radiation in the atmosphere – A hazard to aviation safety? Atmosphere 11(12): 1358. https://doi.org/10.3390/atmos11121358. [CrossRef] [Google Scholar]
- Möller T. 2013. Measurement of the radiation exposure for high altitude flights in the polar region, Ph.D. thesis. Faculty of Mathematics and Natural Sciences. https://macau.uni-kiel.de/receive/diss_mods_00013630. [Google Scholar]
- Moraal H, Belov A, Clem JM. 2000. Design and co-ordination of multi-station international neutron monitor networks. Space Sci Rev 93: 285–303. https://doi.org/10.1023/A:1026504814360. [CrossRef] [Google Scholar]
- Moraal H, Stoker PH. 2010. Long-term neutron monitor observations and the 2009 cosmic ray maximum. J Geophys Res 115: A12109. https://doi.org/10.1029/2010JA015413. [CrossRef] [Google Scholar]
- Oh S, Bieber J, Evenson P, Clem J, Yi Y, Kim Y. 2013. Record neutron monitor rates from galactic cosmic rays. J Geophys Res Space Phys 118: 5431–5436. https://doi.org/10.1002/jgra.50544. [CrossRef] [Google Scholar]
- Poje M, Vukovic B, Radolić V, Miklavčić I, Planinić J. 2015. Neutron radiation measurements on several international flights. Health Phys 108: 344–350. https://doi.org/10.1097/HP.0000000000000192. [CrossRef] [Google Scholar]
- Reames DV. 1999. Solar energetic particles: Is there time to hide? Radiat Meas 30(3): 297–308. [CrossRef] [Google Scholar]
- Ritter B. 2013. The mobile dosimetric telescope: A small active dosimeter for the application in space; development, calibration and measurements, Ph.D. thesis. Faculty of Mathematics and Natural Sciences. https://macau.uni-kiel.de/receive/diss_mods_00013134. [Google Scholar]
- Ritter B, Maršálek K, Berger T, Burmeister S, Reitz G, Heber B. 2014. A small active dosimeter for applications in space. Nucl Instrum Methods Phys Res A 748: 61–69. https://doi.org/10.1016/j.nima.2014.02.030. [CrossRef] [Google Scholar]
- Simpson JA. 2000. The cosmic ray nucleonic component: The invention and scientific uses of the neutron monitor – (Keynote lecture). Space Sci Rev 93: 11–32. https://doi.org/10.1023/A:1026567706183. [CrossRef] [Google Scholar]
- Smart D, Shea M. 2005. A review of geomagnetic cutoff rigidities for earth-orbiting spacecraft. Adv Space Res 36(10): 2012–2020. https://doi.org/10.1016/j.asr.2004.09.015. [CrossRef] [Google Scholar]
- Strauss RD, Potgieter MS. 2014. Is the highest cosmic-ray flux yet to come? Sol Phys 289(8): 3197–3205. https://doi.org/10.1007/s11207-014-0517-7. [CrossRef] [Google Scholar]
- Sullivan J. 1971. Geometric factor and directional response of single and multi-element particle telescopes. Nuc Instrum Methods 95(1): 5–11. https://doi.org/10.1016/0029-554X(71)90033-4. [CrossRef] [Google Scholar]
- Thierfeldt S, Haider C, Hans P, Kaleve M, Neuenfeldt F. 2009. Evaluation of the implementation of radiation protection measures for aircrew in EU member states. Radiat Prot Dosimetry 136(4): 324–328. https://doi.org/10.1093/rpd/ncp170. [CrossRef] [Google Scholar]
- Vainio R, Desorgher L, Heynderickx D, Storini M, Flückiger E, et al. 2009. Dynamics of the earth’s particle radiation environment. Space Sci. Rev. 147(3–4): 187–231. https://doi.org/10.1007/s11214-009-9496-7. [CrossRef] [Google Scholar]
- Wüest M, Evans D, von Steiger R. 2007. Calibration of particle instruments in space physics. ESA SR. International Space Science Institute. https://books.google.co.za/books?id=hmBKNAAACAAJ. [Google Scholar]
Appendix A
The RPiRENA-based active dosimeter
A.1 The detector’s working principle and signal processing
Figure A.1 shows a schematic representation of RPiRENA detector. This particle detector works on the same principles as previously assembled dosimeter prototypes e.g. the Mobile Dosimetric Telescope (Ritter, 2013; Ritter et al., 2014). Here, the detection system draws its power from the 3.3 V and 5 V GPIO (general purpose input output) pins from the connected Raspberry Pi (RPi) zero micro-computer, which is in turn powered via a rechargeable power bank. The instrument draws about 0.4 A of power. Data processing and storage are also performed on the RPi.
Figure A.1 Schematic representation of the RPiRENA detector, divided into an analogue and a digital part. |
The silicone diodes (SDs) are connected to a reversed bias voltage of 45 V to fully deplete the 300 μm thick diodes. When ionising radiation interacts with an SD a short charge pulse is generated due to the separated charges (Leo, 1987). However, the charge pulse is too small to be sensed directly. Therefore, the pulse is first sent to the amplifier section of the analogue electronics. Here, a charge-sensitive amplifier (CSA) amplifies the signals before the resulting pulse is shaped (by the pulse-height shaper, PHS), and then it is digitized by a 16-bit analogue-to-digital converter (ADC). A field programmable gate array (FPGA) is programmed to periodically sample the ADC signals and detect the peak voltage (i.e. the pulse height). Since the pulse height is proportional to the energy deposited, the ADC value is proportional to the energy deposited in the detector after the appropriate calibrations are performed.
A.2 The prototype detector setup
Figure A.2 shows an exterior and interior view of the first RPiRENA-based prototype assembled and tested in 2017. The prototype is placed in a plastic protective housing (shown in the left panel), while the two electronics boards (i.e. the FPGA circuitry and the RPi) along with a power bank and the light-proof SD housing are shown in the middle panel. The right panel shows the two SDs in telescope mode.
Figure A.2 External (left panel) and interior (middle panel) dosimeter housing used for the test flights. The right panel shows a close-up view of silicone diodes (SDs), labelled SD 1 and SD 2, in telescope mode with a separation of 5 mm. |
A.3 Calibration
We follow a similar calibration process as presented by Ritter (2013) and Böttcher (2018). In order to calibrate the RPiRENA ADC output we measure the energy spectra for the following reference sources: Americium-241 (241Am), Caesium-137 (137Cs), Barium-133 (133Ba), and Bismuth-207 (207Bi). The obtained spectra, still in terms of the ADC channel number, are shown in Figure A.3 as binned histograms. For each source, we identify a number of known characteristic photo peaks and Compton edges (indicated by the vertical arrows). The pulse height spectra for 241Am were fitted by a combined exponential curve and Gaussian function and shown as dashed lines. 137Cs, 133Ba, and 207Bi also show a Compton edge and this portion of the distribution was fitted with an additional error function. The value of the Compton edge is also used in the calibrations as shown by Wüest et al. (2007) and Böttcher (2018). The calibration results are listed in Table A.1.
Figure A.3 The pulse height spectra of the calibration sources as measured by the two SDs of the prototype dosimeter. Supplementary data are available on the website. |
Calibration values and statistical comparisons for both SDs.
The energy calibration points are fitted using a linear regression of the form E = (A ± a0) · ADC + (B ± b0), where E is the known energy deposition of the calibration sources and ADC the corresponding ADC channel number. The fit coefficients A, a0, B, and b0 are listed in Table A.2. The results are shown in Figure A.4. The high-gain amplifying stage used here covers energy depositions from as low as the 32 keV and, in principle, this calibration curve can be extrapolated up to 7 MeV in silicon, based on the previous works using the same low energy calibration sources by Ritter (2013). The SD 1 and SD 2 regressions show a slight difference between the SDs, ranging between 0.2% and 5%. This is probably due to small uncertainties in the electronics components and manufacturing tolerances.
Figure A.4 The resulting energy calibration curves for the high-gain channel of both SDs. Supplementary data are available on the website. |
Linear regression coefficients for both SDs.
A.4 Silicon diodes and their geometrical factors
This prototype uses two commercially off-the-shelf rectangular PIN SDs (specifically model s2744-089) manufactured by Hamamatsu Photonics, denoted as SD 1 and SD 2 in Figure A.5 (see side-on view). The width, length, and thickness of each μSD are about 1 cm, 2 cm, and 300 μm, respectively. For ionizing radiation, however, the sensitive effective area of each diode is larger than the quoted active area for photons since energetic particles can penetrate the aluminium housing partially shielding the diodes. For dosimetric purposes the sensitive area is therefore 2.1 cm × 1.1 cm = 2.31 cm2 (Labrenz, 2014). The SDs are arranged in a telescope configuration (with SD 1 facing down, denoted by ⇓, and SD 2 facing up and denoted by ⇑) separated by a distance of 5 mm.
Figure A.5 The s2744-08 Hamamatsu diodes in telescope mode. The top panel shows a side-on view and the SD’s orientation, while the lower panel shows a top view of SD 1. |
A.4.1 Single diode geometric factor
Following the study of Sullivan (1971), the geometry factor (GF) is defined as the product of the effective surface area and the angular coverage of a particle detector. Assuming that the incident particles are isotropic and are interacting with the SD from one side, according to Sullivan (1971), the GF of a single SD is simply given by(A.1)where θ2 is the angle of incidence and is the active effective area of a target material. Therefore, for ionising radiation interacting with the SD used in this work, GF = a · b · π = 13.197 cm2·sr.
A.4.2 Telescope mode geometric factor
For these identical rectangular diodes, separated by 5 mm in telescope mode, the resulting GF is calculated from equation (11) of Sullivan (1971), as 7.257 cm2·sr.
A.5 Effective diode thickness
When isotropic particles interact with both diodes, their trajectory can be projected onto the top diode, as shown in the top right panel of Figure A.5. We are interested in finding the maximum length of this projected tract for every angle 0° ≤ θ ≤ 90°. The longest possible tract is a diagonal line at θc = 26.57° for these specific diodes. We now divide θ into n equally spaced increments and calculate the average length of this maximum tract as(A.2)where θ′ = 90° − θ and i labels all the individual angles considered. For the SDs used in this work, an average maximum projected tract length of 〈x〉 = 15.33 mm is found.
We now switch to a side-on view of the particle telescope as illustrated in the top panel of Figure A.5 and use 〈x〉 as the effective width of the detector. The maximum opening angle of the detector can now be calculated as 2ϕc = 143.86°. The effective thickness of the top diode can now be estimated if we assume an isotropic incident particle distribution from above. For particles entering along the dashed trajectory with an incident angle of ϕ, the top diode will have an effective thickness of leff = 300 μm/cosϕ. The average incident angle of the radiation, for an isotropic distribution, is 〈ϕ〉 = ϕc/2, so that the average effective length becomes(A.3)leading to 〈leff〉 = 370.67 μm.
Appendix B
The CARI model: simulations procedure
The CARI model’s user guide is published by Kyle Copeland10. For the two flights discussed in Section 3, simulations were performed using CARI-7A (ACADEMIC USE, version 4.1.0, published date: October 4, 2019). At aviation altitudes, neutrons and protons dominate the radiation field, while gammas, electrons, positrons, muons, and pions contribute less. Once the CARI model program is started, a command interaction interface will pop-up. To calculate each particle contribution using the CARI-7A model, a simplistic demonstration is given below. For example, to calculate the dose rate and accumulated doses from the neutron component, we followed these options:
-
MAIN MENU ⇒ option 〈2〉 Galactic radiation received on flights
-
FLIGHT MENU ⇒ option 〈2〉 Flight path(s) defined by many waypoints (*.DEG files, same as CARI-6M and -6PM)
-
Enter a file number ⇒ option 〈1〉 JF.10.06.2019Neutron.DEG …file will be imported
-
SELECT RADIATION ⇒ option 〈1〉 …for neutron doses
-
SELECT COSMIC RAY MODEL ⇒ option 〈1〉 GCR: ISO TS15390:2004-MSU-NYMMIK
-
SELECT DOSE TYPE ⇒ option 〈7〉 ABSORBED DOSE IN 0.3 mm Si chip
-
USE THE SUPERPOSITION APPROXIMATION ⇒ option 〈Y,N〉 Y
-
USE DATES FROM FLIGHT DATA FILE ⇒ option 〈Y,N〉 Y
Figure B.1 shows the calculated dose contributions from different radiation sources (colour coded) and the total dose contribution (i.e. the sum of all the individual doses; the black solid line). These calculated absorbed doses in a 0.3 mm Si diode are compared to the RPiRENA measured doses (see the blue and red solid lines representing the SD 1 and SD 2 measurements, respectively).
Figure B.1 Comparison of the CARI-7A model calculated absorbed doses in a 0.3 mm Si diode to the RPiRENA measured doses. The red and blue fluctuating lines represent the SD 1 and SD 2 RPiRENA diode measurements, respectively. Supplementary data are available on the website. |
Cite this article as: Mosotho M, Strauss D, Böttcher S & Diedericks C 2023. The radiation environment over the African continent at aviation altitudes: first results of the RPiRENA-based dosimeter. J. Space Weather Space Clim. 13, 10. https://doi.org/10.1051/swsc/2023007
All Tables
All Figures
Figure 1 Panel A shows the test flight routes from FAJS to EDDF (in blue) while the return flight to FAJS from EDDM is shown in red. Panel B shows the flight altitudes (in terms of atmospheric depth) as a function of the corresponding vertical geomagnetic cutoff rigidity, Pc. Panel A image credit: Google maps (https://sourceforge.net/projects/py-googlemaps/). Supplementary data are available on the website. |
|
In the text |
Figure 2 Panel A shows the measured count rates (SD 1 in blue, SD 2 in red, and coincidence counts in green) during the two tests. Panel B shows the measured absorbed dose rate in each diode, with a minute cadence, while panel C shows the accumulated absorbed dose, . The black curves in the bottom panels show results for the CARI-7A model for 0.3 mm of silicon. The left panels are for the flight from FAJS to EDDF and the right panels are for EDDM to FASJ. Supplementary data are available on the website. |
|
In the text |
Figure 3 LET spectra accumulated for the flight to EDDF (in red) and a return flight to FAJS (in blue). The quality factor, as defined by ICRP103 (2007), is shown in orange. Supplementary data are available on the website. |
|
In the text |
Figure A.1 Schematic representation of the RPiRENA detector, divided into an analogue and a digital part. |
|
In the text |
Figure A.2 External (left panel) and interior (middle panel) dosimeter housing used for the test flights. The right panel shows a close-up view of silicone diodes (SDs), labelled SD 1 and SD 2, in telescope mode with a separation of 5 mm. |
|
In the text |
Figure A.3 The pulse height spectra of the calibration sources as measured by the two SDs of the prototype dosimeter. Supplementary data are available on the website. |
|
In the text |
Figure A.4 The resulting energy calibration curves for the high-gain channel of both SDs. Supplementary data are available on the website. |
|
In the text |
Figure A.5 The s2744-08 Hamamatsu diodes in telescope mode. The top panel shows a side-on view and the SD’s orientation, while the lower panel shows a top view of SD 1. |
|
In the text |
Figure B.1 Comparison of the CARI-7A model calculated absorbed doses in a 0.3 mm Si diode to the RPiRENA measured doses. The red and blue fluctuating lines represent the SD 1 and SD 2 RPiRENA diode measurements, respectively. Supplementary data are available on the website. |
|
In the text |
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.