Open Access

Table 5.

TLD readout, annealing, and calibration procedures applied by the DOSIS/DOSIS 3D participants.

Institute TLD name Readout system Heating Rate [°C/s] Material Preheat Annealing cycle Cooling rate Calibration method Calibration source Glow curve evaluation
DLR TLD-600/-700 TLD-300 Harshaw 5500 (Hamamatsu RC095 HA) 5 LiF:Mg,Ti No preheat 400 °C (1 h), 100 °C (2 h) Slow The same single-chip 137Cs Peak 5 height
CaF2:Tm No preheat 400 °C (1 h), 100 °C (2 h) Slow Peak 5 height
ATI TLD-600/-700 TLD-300 TL-DAT.II (THORN EMI 9635 QB) 5 LiF:Mg,Ti 120 °C (30 min) 400 °C (1 h) Slow The same single-chip 60Co Peak 5 height
CaF2:Tm No preheat 400 °C (1.5 h) Slow Peak 5 height
IFJ (1) MTS-6 MTS-7 MCP-7 RA’94 (THORN EMI 9789 QB) + Harshaw 3500 (ET 9125B) 10 LiF:Mg,Ti 120 °C (30 min) 400 °C (1 h), 100 °C (2 h) Fast Seperate group of TLDs 137Cs Peak integral
LiF:Mg,Cu,P 120 °C (30 min) 240 °C (10 min) Fast Peak integral
SCK·CEN (2) MTS-6 MTS-7 MCP-6 MCP-7 Harshaw 5500 (Hamamatsu RC095 HA) 1 LiF:Mg,Ti No preheat 400 °C (1 h), 100 °C (2 h) Slow Separate group of TLDs from the same batch 60Co Peak integral
LiF:Mg,Cu,P No preheat 240 °C (10 min) Fast Peak integral
MTA EK (3) MTS-6 MTS-7 Harshaw 2000 A-B, PC (Thorn EMI 9235QA) 10 LiF:Mg,Ti No preheat 400 °C (1 h), 100 °C (2 h) Fast The same single-chip 137Cs Peak 5 height
NIRS TLD-100 Harshaw 5500 (Hamamatsu RC095 HA) 25 LiF:Mg,Ti No preheat 400 °C (1 h), 100 °C (2 h) Slow The same batch 137Cs Peak 5 Integral
NASA/SRAG TLD-100 TLD-300 Harshaw 5500 (Hamamatsu RC095 HA) 6 LiF:Mg,Ti 100 °C (30 min) 400 °C (1 h) Slow The same batch 137Cs Peak 5 height
CaF2:Tm 400 °C (1 h), 100 °C (2 h) Fast Peak 5 integral
NPI Al2O3:C CaSO4:Dy RA’94 (THORN EMI 9789 QB) + TOLEDO 654 TLD Reader 10 Al2O3:C No preheat 700 °C (20 min) Fast The same single-chip 137Cs Glow curve integral
CaSO4:Dy 150 °C (22 s) 380 °C (10 min) Fast
(1)

IFJ used RA’94 reader for the DOSIS and Harshaw 3500 reader for the DOSIS 3D experiments.

(2)

SCK·CEN TLD Glow curve evaluation for peak integral: 100 °C–260 °C (MTS) and 170 °C–230 °C (MCP).

(3)

MTA EK started providing TLDs with the start of the DOSIS 3D experiment.

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.

Initial download of the metrics may take a while.