Open Access
Table 6.
OSLD preparation: all groups applied Al2O3:C (Luxel) detectors.
Institute | Readout system | Power [mW/cm2] | Filter | Stimulation Time [s] | Bleaching | Calibration Method | Source | OSL curve evaluation |
---|---|---|---|---|---|---|---|---|
OSU | Risø TL/OSL-DA-15C/D (Green diodes: 525 nm) | 13.3 | 7.5 mm Hoya-U340 | 600 | Yellow light over night | Single “chip” | 90Sr/90Y (calib. against 60Co) | (2) |
SCK·CEN | Homemade (Ar+ laser: 488 nm) | 120 | Melles Griot 01MCG019 | 100 | No (daylight) | Separate group from the same batch | 60Co | |
NASA/SRAG | Risø TL/OSL-DA-15C/D (Green diodes: 525 nm) | 40 | 7.5 mm Hoya-U340 | 300 | Bleached with 525 nm for 120 s | Same batch | 137Cs | (3) |
NIRS | MicroStar InLight (Landauer Inc.) (Green diodes: 525 nm) | 173 (1) | 7.5 mm Hoya-B390 | 1 | Fluorescent light without UV | Same batch | 137Cs |